Alkalinity (Gran titration)
How the titration runs, the DIC-corrected Gran fit, and what the quality flags mean.
The run
- Rinse Cell A twice with tank water; home the acid pump if needed; fill and level to 20.0 mL; stir 15 s; read temperature and starting pH.
- Initial dose. 97 % of the acid expected from the previous result (or
initial_dose_ulon the first run) in one shot, so the titration lands just short of equivalence. - Approach. 200 µL steps until pH < 4.3.
- Titrate. 60 µL steps, waiting for pH drift < 0.005 pH/s after each (timeout 25 s), until pH < 3.0 or 14 points have been collected in the 3.0–3.8 window.
- Fit, store, drain, rinse once, park the cell full of tank water.
A typical 8 dKH titration uses 2.9 mL of 0.02 N acid and takes six to eight minutes.
The math
Below pH 4 the carbonate species are essentially all CO₂, and the charge balance on the solution is
Multiplying by the total volume (V₀ + v) and substituting TA_remaining·(V₀+v) = TA₀·V₀ − c·v and C_T·(V₀+v) = C_T0·V₀ gives a straight line:
A least-squares line of the left side against x = V₀K₁/[H⁺] has intercept −TA₀V₀ and slope C_T0: alkalinity and an estimate of dissolved inorganic carbon from the same points. With pK₁ = 5.86 (seawater, 25 °C), the DIC term is 2–4 % of [H⁺] in the window and ignoring it — as the classic Gran extrapolation does — biases alkalinity 2–5 % low. OpenReef reports both: ta_meq_l (corrected, primary) and ta_classic_meq_l.
dKH = 2.80 × TA (meq/L); 1 meq/L = 50 ppm CaCO₃.
Why it tolerates a mediocre electrode
A constant pH offset changes every [H⁺] by the same factor; the fit’s x-intercept does not move. Slope errors matter more, so calibrate at pH 7 and 4 (bracketing the window) every two weeks, and let the EZO circuit’s temperature compensation run from the DS18B20.
Standardising the acid
Rather than trusting the label, run the titration on a reference of known alkalinity — 0.1000 g/L anhydrous Na₂CO₃ (dried at 250 °C) is 1.887 meq/L = 5.28 dKH — and solve for the normality that reproduces it (POST /api/v1/calibration/alkalinity/standardize). The DIC-corrected fit depends on c, so the solver bisects on c until TA matches. Re-standardise whenever you refill the acid bottle.
Quality flags
| Flag | Meaning | What to do |
|---|---|---|
low_r2 |
fit R² < 0.999 | electrode noisy or stirring poor; check bubbles on the bulb |
noisy_fit |
relative residual rms > 1 % | same; halves the quality score |
v_eq_outside_expected |
equivalence volume not before the first window point | first dose too large; lower initial_dose_ul |
dic_estimate_implausible |
DIC < 0 or > 2× TA | window too narrow or pH scale off; recalibrate pH |
The quality score (0–1) is derived from R² and residuals; the dashboard greys out points below 0.5.
Configuration
chemistry:
alkalinity:
acid_normality: 0.0200 # replaced by standardisation
sample_ml: 20.0
initial_dose_ul: 900
step_ul: 60
gran_ph_min: 3.0
gran_ph_max: 3.8
min_points: 6
max_points: 14
stable_dph_per_s: 0.005
stable_timeout_s: 25