Introduction

Overview

What OpenReef is, what it measures, and how the parts fit together.

OpenReef is a bench-top analyzer for reef aquaria. It measures total alkalinity by Gran acid titration and nitrate and phosphate by self-blanked colorimetry, on a schedule, and publishes every result through a REST and WebSocket API. You build it from off-the-shelf parts and a few printed pieces for about $849.

Parameter Method Range Resolution Target accuracy Sample Cycle
Alkalinity Gran titration, 0.02 N HCl, pH electrode 4–14 dKH 0.02 dKH ±0.05 dKH 20 mL 6–8 min
Phosphate (PO₄) Molybdenum blue, 850/880 nm 0.00–1.00 ppm 0.01 ppm ±0.02 ppm or 5 % 10 mL 10–12 min
Nitrate (NO₃) Reduction + Griess azo dye, 540 nm 0–50 ppm 0.1 ppm ±1 ppm or 5 % 10 mL 15–30 min

What is in the box

 ┌──────────────────────────────┬──────────────┐
 │  WET BAY                     │  DRY BAY     │
 │  Cell A  20 mL titration     │  Raspberry   │
 │     pH electrode, DS18B20,   │  Pi 4        │
 │     stir bar                 │              │
 │  Cell B  10 mL cuvette       │  5× TMC2209  │
 │     AS7341 + white/IR LEDs   │  MCP23017    │
 │  5× syringe pumps            │  8× MOSFET   │
 │     acid, NP R1/R2/R3a/R3b   │  relay       │
 │  6× peristaltic (fill/level/ │              │
 │     drain per cell), 2 stir  │  12 V 5 A    │
 │  reagent tray · leak sensor  │              │
 └──────────────────────────────┴──────────────┘
                 waste bottle (HDPE, float switch)

One Python service (firmware/openreef_analyzer) runs the fluidics, the chemistry math, a scheduler, SQLite history, the dashboard and the API. The same code runs against a hardware simulator with a carbonate-system model, so you can develop and test without the analyzer.

Design goals

  1. Trustworthy numbers. Every result stores its raw titration curve or absorbance channels, the calibration it used and a quality score, so a bad reading is visible instead of silently logged.
  2. Off-the-shelf parts only, total under $1000 including the first two months of reagents.
  3. One codebase, one board. No microcontroller firmware to flash; the Pi drives steppers with DMA-timed pulses.
  4. Reagents you can buy. Neptune’s Trident NP kit for NO₃/PO₄ and ordinary 0.1 N HCl for alkalinity, with lab-standard chemistry documented as a fallback.
  5. Fail safe. Leak sensor, waste-full float, reagent accounting, homing before every fill and a watchdog stop pumps before they make a mess.

Default schedule and consumption

Alkalinity every 6 hours; nitrate and phosphate once a day. That uses about 75 mL of tank water and 1.5 mL of 0.02 N acid per alkalinity test, one Trident NP reagent set every 60 days, and 500 mL of 0.1 N HCl every 4 months.

Where to go next