Everything hardware- and chemistry-specific lives in one YAML file.

The service reads $OPENREEF_CONFIG, then ./config.yaml, then /etc/openreef/config.yaml, then falls back to config.example.yaml. Validate with openreef check-config.

Sections

Section What it holds
hardware backend (sim/rpi), I²C addresses, Pi GPIO pins, the five steppers (STEP/DIR pins, home pin, lead-screw and syringe geometry, reagent, target cell), the eight dc channels (MCP23017 pin, flow rate, cell, role), the two cells (metered volume, path length)
chemistry reagents (bottle sizes and pump assignment), the alkalinity titration profile, the phosphate and nitrate colorimetric profiles (doses, mixing, development, LED and channels, temperature coefficient, starting calibration coefficients)
schedule list of {test, cron, enabled}; editable live through the API
limits [low, high] alert bands per parameter
hydros, mqtt, webhooks integrations
api bind address, port, API keys, CORS
sim_speedup simulator time factor

Pin numbering

step_pin, dir_pin, leak_pin, alarm_relay_pin, led_*_pin, heater_pwm_pin are Raspberry Pi BCM numbers. Stepper home_pin and DC pin are MCP23017 pins: 0–7 = port A (inputs), 8–15 = port B (outputs).

Syringe geometry

steppers:
  - { name: SP1, step_pin: 27, dir_pin: 22, home_pin: 0, reagent: hcl, cell: A,
      steps_per_mm: 1600, syringe_bore_mm: 12.0, stroke_mm: 45, max_speed_mm_s: 2.5 }

steps_per_mm = 200 × microsteps / lead_mm. For a 1 mm-lead T8×1 screw use 3200; for a 10 mL syringe (bore ≈ 14.5 mm) set syringe_bore_mm accordingly. The software computes µL per step from these.

Values that persist outside the file

Calibration results (acid_normality, curve coefficients, secondary ratios), reagent remaining volumes, edited limits and schedule are stored in the SQLite settings and reagents tables and override the YAML. Delete the database to start clean.

Example

The full annotated example is firmware/openreef_analyzer/config.example.yaml in the repository.